FMA Vacuum Insertion Tool
| Vacuum Insertion Tool - Model VPI-002 |
The model VPI-002 FMA Vauum Insertion Tool has been designed to allow straight forward and
easy insertion of the FMA into tissue. It can be used with any commercial micro-drive
system. The finely ground 16 gauge stainless steel cannula used to pick up and hold the FMA
substrate can be mounted to most microelectrode holders or micro-drive electrode assembly
clamping devises. The FMA insertion tool utilized a vacuum pump system designed to provide
sufficient suction in order to allow slight manipulations of the FMA substrate around the
tissue insertion site.
It is recommended that the FMA be inserted very slowly in order to minimize dimpling of the
cortex. It incorporates a "wand" that will allow the FMA substrate to be released without
having to turn off the vacuum pump. A stand is provided to hold the wand near the surgical
field. The wand and stand may be gas sterilized along with the Tygon tubing leading from
the wand to the stainless steel cannula, which is used to pick up the FMA substrate. The
Tygon tubing and stainless steel cannula may be autoclaved if that is preferred. However,
it is not recommended to autoclave the wand.
| Complete System: |
Specifications: |
 |
| Model |
FV-60-110 |
| Flow Rate |
2.51/mind (0.088 cfm) |
| Ultimate Pressure |
-600 mmHg (-80 kPa) |
| Power Requirement |
100-120 VAC |
| Consumption Power |
10.0 W |
| Dimensions (D x W x H) |
137 x 88 x 133 mm |
| Weight |
1250 g |
| Duty Cycle |
Continuous |
| Life Expectancy |
> 8760 hours |
|